What samples have you tested?
This kit has been tested and validated in urine, plasma, and serum samples.Can I store unused reagents for future use?
Yes, unused reagents can be stored according to the assay protocol.How do you use this kit on deproteinated plasma or serum samples with an internal standard procedure?
Samples will need three separate reactions: 1) sample plus standard2) sample alone 3) sample blank. For the internal standard prepare 200 µL 1250 µM standard by mixing 125 µL 2 mM Standard and 75 µL dH2O. For the sample plus standard well, add 5 µL 1250 µM standard and 20 µL sample. For the sample and sample blank wells, add 5 µL dH2O and 20 µL sample. Prepare enough Working Reagent (WR) for all assay wells by mixing, for each well, 80 µL Assay Buffer, 1 µL Enzyme A, 1 µL Enzyme Mix, 1 µL ATP, 1 µL Dye Reagent. Prepare a WR without the Enzyme A for the sample blank well(s). Add 80 µL of the WR to each well. Tap plate briefly to mix. Incubate at room temperature for 60 min. Use a plate reader to read OD570nm. The sample creatinine concentration is computed as follows: [Creatinine] = (RSAMPLE – RBLANK)/(RSTANDARD – RSAMPLE) x [Standard]/4 x n(µM)where RSAMPLE, RBLANK, and RSTANDARD are the OD or fluorescence values of the Sample, Sample Blank, and the Sample plus Standard respectively. n is the sample dilution factor. Note: The volume of the internal standard is 4× lower than the sample volume; thus, the internal standard concentration should be divided by 4.For more detailed product information and questions, please feel free to Contact Us. Or for more general information regarding our assays, please refer to our General Questions.