The blank value is higher than the standard values. What is wrong?
This is completely normal. The assay is measuring the decrease in absorbance at 500nm. The higher the concentration of biotin, the lower the absorbance will be in the calculation.My sample is reporting much lower levels of biotin than I was expecting?
The biotin levels in your sample may be significantly greater than the kit detection limit. We recommend diluting your sample and assaying again. You may wish to perform a serial dilution on the sample to determine the optimal dilution factor. Aim for a dilution that yields a concentration read out somewhere in the middle of the standard curve. Correct for the dilution factor in the calculation.Some of my samples are getting negative concentration read outs?
This is likely indicative no biotin or biotin below the kit detection limit.The kit arrived thawed and at room temperature, but the protocol says to store it at -20°C. Is the kit still ok to use?
Yes, the kit is shipped at room temperature and is stable for the duration of shipment at that temperature. Upon receipt, the kit should be stored at -20°C.For more detailed product information and questions, please feel free to Contact Us. Or for more general information regarding our assays, please refer to our General Questions.