What samples have you tested?
The strips have been tested on red wine, white wine, homogenized whole milk, cultured buttermilk, almond milk, yogurt, beer, apple juice, lemonade, grape juice, orange juice, cranberry juice, lime and lemon juice.My sample turns the strip very dark purple, how can I determine the D-lactate concentration?
Very dark purple indicates that the diluted sample concentration is greater than 500 mg/L. To obtain a more accurate concentration, the sample should be further diluted and retested. For example, if the sample was diluted 5×, try diluting 21×.I don’t have access to a pipetteman. How can I accurately measure out my samples?
We offer exact volume transfer pipettes as an accessory. For samples requiring a 21× dilution, order the 20 µL transfer pipettes (TP20). For samples requiring a 5× dilution, order the 100 µL transfer pipettes (TP100).Can I store unused reagents for future use?
Yes, unused reagents can be stored according to the assay protocol. The strips should be kept in a dry, cool location and protected from light.For more detailed product information and questions, please feel free to Contact Us. Or for more general information regarding our assays, please refer to our General Questions.