The background is very high. Is this normal?
The background absorbance is typically ~ 0.5. This is completely normal.Can I store the diluted standards (250 µM, 150 µM, 75 µM, and 0 µM) to use for later?
We recommend preparing the standards fresh from the 50 mM each day.Do I need to run a standard curve with every assay?
Yes.How should I store the kit?
The kit is shipped at room temperature. Upon delivery, store at -20°C.Precipitate formed when I added the reagent to the sample. What should I do?
The Reagent is acidic and may cause precipitation in samples that have significant amounts of protein. You may need to dilute your sample, repeat the assay, and correct for the dilution factor in the calculation.What is the Carbonyl Standard?
The standard is pyruvate, which has two carbonyl groups per molecule. So the 50 mM Carbonyl Standard is actually 25 mM pyruvate.My sample has high levels of nucleic acids. Will this interfere?
Nucleic acids can artificially increase the carbonyl count if they are present in high levels. If the ratio of 280/260 nm is less than 1, samples should be mixed with 1% streptomycin sulfate to denature the nucleic acids. Mix well, incubate the samples at room temperature for 15 minutes and then centrifuge at 5000xg for 10 minutes. Use the supernatant in the assay.For more detailed product information and questions, please feel free to Contact Us. Or for more general information regarding our assays, please refer to our General Questions.